SYSTem Subsystem 12
N1911A/1912A P-Series Power Meters Programming Guide 531
[SENS[1]]|SENS2:CORR:CSET[1]|
CSET2[:SEL]
not affected Selected sensor calibration table
[SENS[1]]|SENS2:CORR:CSET[1]|
CSET2:STAT
not affected Sensor calibration table disabled
[SENS[1]]|SENS2:CORR:DCYC|GAIN3
[:INP][:MAGN]
1.000 % Duty cycle factor
[SENS[1]]|SENS2:CORR:DCYC|GAIN3:ST
AT
OFF Duty cycle correction disabled
[SENS[1]]|SENS2:CORR:FDOF|GAIN4[:I
NP][:MAGN]
not affected Return frequency dependent offset
[SENS[1]]|SENS2:CORR:GAIN2:STAT OFF Channel offset disabled
[SENS[1]]|SENS2:CORR:GAIN2:STAT
[:INPut][:MAGNitude]
0.0 dB Enter channel offset value
[SENS[1]]|SENS2:DET:FUNC NORM Measurement mode
[SENS[1]]|SENS2:FREQ[:CW|:FIX] +50.000 MHz Frequency setting
[SENSe[1]]|SENS2:MRAT NORM Measurement speed
[SENS[1]]|SENS2:POW:AC:RANG upper Upper range selected
[SENS[1]]|SENS2:POW:AC:RANG:
AUTO
ON Auto-ranging selected
[SENS[1]]|SENS2:SPE 20 readings/
second
Speed
[SENS[1]]|SENS2:SWE[1]|2|3|4
:OFFS:TIME
0 Set delay
[SENS[1]]|SENS2:SWE[1]|2|3|4
:TIME
Gate 1: 100 µs
Other gates: 0 sec
Set time gated period
[SENS[1]]|SENS2:TRACe:OFFSet:
TIME
0 Delay
[SENS[1]]|SENS2:TRACe:TIME 100 µs Duration of trace
[SENS[1]]|SENS2:V2P ATYP Select linearity correction
SYST:GPIB[:SELF]ADDR not affected Power meter address
TRAC[1]|2:STAT OFF Disable trace capture
TRAC[1]|2:UNIT dBm Trace units
TRIG[1]|2:DEL:AUTO ON Insert settling time delay
Command Setting Comments