Instructions for use Zeus Infinity Empowered SW 2.n 129
Operation
PEEP modifications
Depending on configuration in the basic settings,
the inspiration pressure Pinsp can be automatically
changed when modifying the PEEP value,
see page 349.
For dependent setting of the change in Pinsp:
1 Touch the PEEP therapy control (I).
The color turns yellow. The value of ventilation pa-
rameter P
insp (J) also turns yellow.
2 Set PEEP and confirm with the rotary knob.
The value for P
insp is automatically adjusted.