Monitoring
222
Instructions for use Zeus Infinity Empowered SW 2.n
Starting the NIBP measurement
z In the Sensors/Parameters dialog window,
touch the NIBP start button (B).
Or
z In the main menu bar, touch the NIBP start (J)
button.
The NIBP measurement is performed.
Measuring intervals
The measurements are taken at the intervals set
under NIBP mode (B).
Exception: All intervals which are a multiple of
5 minutes synchronize from the second measure-
ment to 5-minute steps of the current time so that
they are in the measuring grid of the trend display.
In the Venous stasis mode, the cuff is inflated for
up to 2 minutes. Pressure values, see page 223.
In the Cont. 5 Min. mode, single measurements
are continuously taken over 5 minutes.
Parameter boxes of NIBP measurements
Single measurements (Single mode)
Start and stop single measurement:
z Touch the NIBP start or NIBP stop button.
301
1:35
J
216
145/82
160
90
(105)
118/81
160
90
18
118/81
160
90
(94)
118/81
160
90
(94)
118/81
160
90
86
5 min
S
Cont.
S
NIBP
NIBP
NIBP
NIBP
NIBP
mmHg
mmHg
mmHg
mmHg
mmHg
10:12
Cuff
S
10:08
10:12
Cuff
S
Stasis
S
10:12
15 min
10:12
Single measurement
Continuous measurement
Stasis measurement
Auto measurement (5 min)
Auto measurement (15 min)