Monitoring
240
Instructions for use Zeus Infinity Empowered SW 2.n
Preparing for NMT
1 Attach the black stimulation electrode (A) and
the red stimulation electrode (B) to the arm
(above the nervus ulnaris).
2 Connect the pod connecting cable (C) to the Tri-
dent pod and to the Aux/Hemo 2/3 connector
on Zeus IE.
3 Attach the NMT sensor (D) to the thumb and in-
dex finger. Secure with a rubber band.
4 Connect the electrodes and NMT sensor to the
NMT accessory cable (E).
5 Connect the NMT accessory cable to the Tri-
dent pod (F).
WARNING
Risk of burns
When fitting the module, make sure there is
sufficient air ventilation/heat dissipation and
that there can be no direct contact between
the patient and the exterior of the module.
WARNING
Make sure that there is no contact between the
electrodes and other conducting parts.
Cover stimulation electrodes with insulating
material so that, e.g., catheters are not sub-
jected to stimulation.