Instructions for use Zeus Infinity Empowered SW 2.n 263
Fault – Cause – Remedy
Warning 150 Battery very low The capacity of the uninter-
ruptible power supply is ex-
hausted (<10 %).
Restore power supply. Pre-
pare for man. ventilation.
Minimize power consump-
tion. Check battery capacity
display. If capacity is <10 %
the device may be switched
off!
Caution 100 BIS high BIS value above higher
alarm limit. Patients hypnotic
state not appropriate.
Check patient. Treat, if nec-
essary. Increase agent
dose if necessary. Change
alarm limit as appropriate.
Caution 100 BIS low BIS value below lower alarm
limit.
Check patient. Treat, if nec-
essary. Decrease agent
dose if necessary. Change
alarm limit as appropriate.
Caution 050 BIS out of range BIS parameter value out of
range.
Check patient and treat, if
necessary.
Caution 050 BIS out of range BIS parameter value out of
range.
Check patient and treat, if
necessary.
Note 210 BIS signal quality
poor
SQI (Signal Quality Index)
value goes below 50 %.
Check sensor. This may oc-
cur as a result of artefact
caused e.g. by eyeblinks or
motion. Isolate patient from
auxiliary equipment. Isolate
patient from extraneous
movement or vibration.
Check patient. Treat, if nec-
essary.
Note 050 BIS/NMT alarms
off
BIS and NMT alarms sus-
pended
Watch measured values to
monitor patients status.
Classifi-
cation
Pri-
ority
Alarm text Cause Remedy