Fault – Cause – Remedy
264
Instructions for use Zeus Infinity Empowered SW 2.n
Note 100 BISX impedance
check failed
BISx pod failed the sensor
impedance check.
Retry impedance check.
Re-prep the electrodes and
retry. Replace sensor, if
necessary and perform
manual impedance check.
Note 100 BISX impedance
check in progress
BISx pod is performing sen-
sor impedance check.
Wait until impedance check
is completed. Cancel check
and retry.
Note 100 BISX pod discon-
nected
BISx-Pod not connected. Check cables and connec-
tions. Exchange BISx-pod,
if necessary.
Note 105 BISX pod failed BISx pod self test failed.
BISx pod hardware failure.
BISx pod software failure.
Check BISx pod connec-
tion. Call your biomed serv-
ice or DrägerService.
Note 020 BISX pod initialis-
ing
BISx pod is performing star-
tup test.
Wait until test is completed.
Note 100 BISX replace sen-
sor
Unknown sensor type. Too
many uses./ Sensor worn
out. Invalid sensor.
Remove unknown/worn
out/invalid sensor and re-
place with new/compatible
sensor.
Note 100 BISX sensor dis-
connected
Sensor malfunctioning or
disconnected.
Check cables and connec-
tions. Replace sensor, if
necessary.
Note 100 BISx sensor has
expired
Unknown sensor type. Too
many uses./ Sensor worn
out. Invalid sensor.
Remove unknown/worn
out/invalid sensor and re-
place with new/compatible
sensor.
Note 100 BISx sensor will ex-
pire after this ses-
sion
The usage limit for the BISx
SRS sensor has been
reached.
Prepare a new BISx SRS
sensor for next application.
Note 250 Bolus 1 rate high Pump 1 Bolus Rate Limit ex-
ceeded
Check the bolus rate of the
pump. Adjust settings if
necessary.
Classifi-
cation
Pri-
ority
Alarm text Cause Remedy