Instructions for use Zeus Infinity Empowered SW 2.n 265
Fault – Cause – Remedy
Note 250 Bolus 2 rate high Pump 2 Bolus Rate Limit ex-
ceeded
Check the bolus rate of the
pump. Adjust settings if
necessary.
Note 250 Bolus 3 rate high Pump 3 Bolus Rate Limit ex-
ceeded
Check the bolus rate of the
pump. Adjust settings if
necessary.
Note 250 Bolus 4 rate high Pump 4 Bolus Rate Limit ex-
ceeded
Check the bolus rate of the
pump. Adjust settings if
necessary.
Note 210 Breathing system
not locked
The breathing system is not
properly locked.
Insert the breathing system
and lock it properly.
Note 100 C.O. blood temp.
high
Blood temperature is outside
alarm limits: A physiological
condition. Inappropriate
alarm limits. A defective sen-
sor or cartridge.
Check the patient and take
appropriate action. Change
alarm limits. Check equip-
ment and replace if neces-
sary.
Note 100 C.O. blood temp.
low
Blood temperature is outside
alarm limits: A physiological
condition. Inappropriate
alarm limits. A defective sen-
sor or cartridge.
Check the patient and take
appropriate action. Change
alarm limits. Check equip-
ment and replace if neces-
sary.
Note 100 C.O. check injec-
tate probe
Injectate temperature probe
not connected during a
measurement.
Connect the probe and re-
peat the measurement.
Note 100 C.O. complete The C.O. average has been
saved.
Nothing required.
Note 100 C.O. hardware fail C.O. measurement system
failed.
Exchange components.
Call DrägerService.
Note 100 C.O. injectate temp
= 0°C?
No injectate probe is con-
nected. Monitor assumes
temperature of 0 °C. Refer-
ence resistance too high or
low. Unknown catheter type.
Use either 0 °C injectate or
temp. probe. Repeat meas-
urement. Check catheter
and reconnect pod. Call
DrägerService.
Classifi-
cation
Pri-
ority
Alarm text Cause Remedy