Fault – Cause – Remedy
266
Instructions for use Zeus Infinity Empowered SW 2.n
Note 100 C.O. no temp
change
Temperature change was
<0.1°, because: C.O. started
without injecting. Injected
volume too small or too
warm. Defective catheter.
Repeat the measurement.
Use a larger injectate vol-
ume. Use colder injectate.
Replace the catheter.
Note 200 C.O. out of range C.O. out of range due to: A
physiological condition. Un-
stable baseline, incorrect inj.
volume, catheter size, comp.
constant. Defective catheter,
cable, or cartridge.
Check the patient and take
appropriate action. Use
cooler injectate. Enter the
correct values in C.O.
menu. Repeat the meas-
urement.
Note 100 C.O. transducer fail C.O. transducer failed. Check and reconnect C.O.
transducer. Replace com-
ponent. Call DrägerService.
Note 100 C.O. unplugged A cable or transducer has
been disconnected.
Reconnect the cable or
transducer. If message per-
sists, replace defective part.
Note 100 C.O. unstable
baseline
Temperature curve did not
return to baseline within
30 sec. after C.O.-start Un-
stable patient temperature.
Defective catheter, cable,
cartridge.
Follow hospital procedures.
Repeat the measurement.
If message persists, re-
place the defective compo-
nents.
Note 100 C.O. use cooler in-
jectate
Injectate is warmer than
+27 °C. Temp.difference
<5 °C. Injectate probe not
connected. Defective cable
or hemodynamic pod.
Use a cooler injectate within
–5° to +27 °C. Check probe
connection. Repeat the
measurement. Replace the
probe.
Note 050 Cardiac bypass
mode
All cardiovascular alarms
and apnea alarms are sus-
pended. Use only during
Cardiac Bypass operation.
Please pay special atten-
tion! After manoeuvre or
Cardiac Bypass operation
reactivate alarms immedi-
ately.
Classifi-
cation
Pri-
ority
Alarm text Cause Remedy