Instructions for use Zeus Infinity Empowered SW 2.n 267
Fault – Cause – Remedy
Caution 100 Check NIBP cuff
size
Pulsations are too small to
determine blood pressure.
Check the cuff size. Move
the cuff to another limb.
Caution 010 Check patient cate-
gory
Patient age, height or weight
do not match the values that
are typical for the selected
patient category.
Check patient category and
patient data. Change pa-
tient category or patient
data, if necessary. This
alarm be confirmed by
pressing the Audio paused
key.
Note 000 CO
2 measurement
not accurate
Measured CO2-values out of
specified accuracy. Calibra-
tion not possible.
If message occures repeat-
edly call DrägerService.
Note 200 CO
2 sensor failure Failure of CO2 sensor. Monitor ventilation by
minute volume. Call
DrägerService.
Caution 010 CO
2 water
tr./samp. line?
CO2 sample line blocked.
CO2 water trap or gas meas-
urement system blocked.
CO
2 water trap full or ex-
pired. Gas measurement
and auto delivery not possi-
ble.
Check CO
2 sample line and
CO2 water trap. Check filter
at the Y-piece and gas
measurement system. Ex-
change, if necessary. Ex-
change water trap at least
every 4 weeks.
Note 255 CO
2 water trap or
sample line
blocked
CO2 sample line blocked.
CO2 water trap or gas meas-
urement system blocked.
CO
2 water trap full or ex-
pired.
Check CO
2 sample line and
CO2 water trap. Check filter
at the Y-piece and the gas
measurement system. Ex-
change, if necessary. Ex-
change water trap at least
every 4 weeks.
Warning 150 Cont. AW pressure Airway Pressure above
(auto-) limit for more than
15 s.
Check airway. Release
pressure!
Classifi-
cation
Pri-
ority
Alarm text Cause Remedy