Fault – Cause – Remedy
268
Instructions for use Zeus Infinity Empowered SW 2.n
Caution 100 CPP M high Invasive pressure value out-
side alarm limits.
Check patient and take ap-
propriate action. Change
the alarm limits if required.
Caution 090 CPP M low Invasive pressure value out-
side alarm limits.
Check patient and take ap-
propriate action. Change
the alarm limits if required.
Note 050 CV alarms off All cardiovascular alarms
suspended.
Please pay special atten-
tion! To be used only during
startup. After manoeuvre
reactivate alarms immedi-
ately!
Note 050 CVP check zero Pressure transducer for
specified parameter requires
zeroing. The blood pressure
zero of monitor may not cor-
respond to the peripheral de-
vice.
Zero the transducer.
Note 025 CVP hardware fail Invasive blood pressure
hardware failure.
Check hardware and re-
place. Call DrägerService.
Caution 090 CVP M high Invasive pressure value out-
side alarm limits.
Check patient and take ap-
propriate action. Change
the alarm limits if required.
Check equipment and re-
place if necessary.
Caution 100 CVP M low Invasive pressure value out-
side alarm limits.
Check patient and take ap-
propriate action. Change
the alarm limits if required.
Check equipment and re-
place if necessary.
Note 150 CVP nonstatic Press. transducer zeroing
failed because of: Excessive
signal noise A non-static
waveform
Keep all tubing motionless,
then rezero. Change the
transducer. Check 3-way
tap, then rezero.
Classifi-
cation
Pri-
ority
Alarm text Cause Remedy