Instructions for use Zeus Infinity Empowered SW 2.n 269
Fault – Cause – Remedy
Caution 050 CVP out of range Invasive pressure signal out
of measurement range.
Check patient and take ap-
propriate action. Access the
Pressure Labels menu and
assign correct label. Check
equipment and replace if
necessary.
Note 050 CVP transducer fail Pressure transducer failed.
Invasive pressure cannot be
measured.
Check transducer and con-
nections. Exchange trans-
ducer.
Note 025 CVP transducer
unplugged
Pressure transducer for
specified parameter is either
unplugged or defective.
Active pressure: reconnect
or replace the cable. Inac-
tive pressure: turn off
alarms.
Note 050 CVP zeroing Mercury calibration in
progress.
Complete calibration before
you begin monitoring the
patient.
Note 255 Deactivate CBM
mode
Cardiovascular activity de-
tected. Device is in cardiac
bypass mode.
Deactivate cardiac bypass
mode!
Caution 100 DIVA Air supply
pres. low
Pressure supply of vol. agent
delivery failed. Leakage at
pressure connection of mod-
ule. Failure of delivery mod-
ule.
Check central Air supply!
Check connection of the
module. Exchange the de-
livery module.
Caution 070 DIVA module 1
empty
No volatile agent in DIVA
module 1.
Fill up the module immedi-
ately. Before opening the
lock: ventilate reservoir. Fill-
ing is possible during oper-
ation.
Warning 030 DIVA module 1 fail-
ure
DIVA module 1 or port 1
failed. Or DIVA module 1
was terminated by protection
system.
Use alternative port. Ex-
change module.
Classifi-
cation
Pri-
ority
Alarm text Cause Remedy