Fault – Cause – Remedy
270
Instructions for use Zeus Infinity Empowered SW 2.n
Caution 110 DIVA module 1 in-
compat.
Software of DIVA module 1 is
not compatible. Module can-
not be used.
Use module with compati-
ble software.
Call DrägerService.
Note 050 DIVA module 1
level low
Filling level of DIVA module 1
<50 mL.
Fill up the module as soon
as possible. Filling is possi-
ble during operation.
Caution 070 DIVA module 2
empty
No vol. agent in DIVA mod-
ule 2
Fill the module. Before
opening the lock ventilate
reservoir
Warning 030 DIVA module 2 fail-
ure
DIVA module 2 or port 2
failed. Or DIVA module 2
was terminated by protection
system.
Use alternative port. Ex-
change the module.
Caution 110 DIVA module 2 in-
compat.
Software of DIVA module 2 is
not compatible Module can-
not be used.
Use module with compati-
ble software.
Call DrägerService.
Note 050 DIVA module 2
level low
Filling level of DIVA module 2
<50 mL.
Fill up the module as soon
as possible. Filling is possi-
ble during operation.
Note 255 DIVA module war-
mup
DIVA module is being
warmed up. Delivery will
start after approx. 2 min.!
Wait until the warm up proc-
ess is finished.
Caution 000 DIVA off! Module
removed
DIVA delivery interrupted!
Module removed. In Auto
control mode agent will be
washed out quickly.
Use fresh gas mode to stop
wash out. Insert DIVA mod-
ule. Use 2nd DIVA module.
Caution 010 DIVA protect fail DIVA protection system
failed. Protection system is
monitoring with extended
alarm limits.
Use mode Fresh-gas con-
trol. Set fresh-gas flow
higher than 3 L/min. NOTE:
At lower flows DIVA will be
terminated.
Call DrägerService.
Classifi-
cation
Pri-
ority
Alarm text Cause Remedy