Instructions for use Zeus Infinity Empowered SW 2.n 271
Fault – Cause – Remedy
Warning 045 DIVA+mixer cutoff DIVA and mixer failure. DIVA
and mixer delivery inter-
rupted.
Open O2 Safety valve. Con-
tinue anesthesia with IV
drug. Call DrägerService.
Warning 045 DIVA+mixer failure DIVA and mixer failed. Deliv-
ery not possible.
Open safety O
2 supply.
Continue anesthesia with
IV drug.
Note 120 Drug data read er-
ror
Error while reading drug data
from USB stick or internal
database.
Check and reinsert the USB
stick. Restart import proce-
dure. Check drug library.
Note 120 Drug data write er-
ror
Error while writing drug data
to USB stick or internal data-
base. USB stick full or write
protected.
Check and reinsert the USB
stick. Insert a USB stick that
has enough space. Re-
move write protection.
Check drug library.
Note 000 ECG hardware fail-
ure
ECG hardware failure de-
tected by 1 mV test at star-
tup.
Call DrägerService.
Note 000 ECG/Baseline arte-
fact
ECG beats are disturbed.
Patient movements, shivers,
tremors. Bad electrode con-
tact. Excessive signal noise
interference from auxiliary
equipment. Artefact is block-
ing arrhythmia classification.
Check electrode applica-
tion. Isolate patient from
auxiliary equipment.
Caution 020 Empty syringe 1
mode
Pump 1 is running in 'Empty
syringe mode'.
Replace syringe 1, if neces-
sary.
Caution 020 Empty syringe 2
mode
Pump 2 is running in 'Empty
syringe mode'.
Replace syringe 2, if neces-
sary.
Caution 020 Empty syringe 3
mode
Pump 3 is running in 'Empty
syringe mode'.
Replace syringe 3, if neces-
sary.
Caution 020 Empty syringe 4
mode
Pump 4 is running in 'Empty
syringe mode'.
Replace syringe 4, if neces-
sary.
Classifi-
cation
Pri-
ority
Alarm text Cause Remedy