Instructions for use Zeus Infinity Empowered SW 2.n 273
Fault – Cause – Remedy
Warning 255 Fresh gas in-
tern/extern?
Failed to switch between ex-
ternal and internal fresh gas
outlet. No fresh gas is deliv-
ered to the requested
breathing system!!!
Check fresh gas flow and
anesthetic delivery. Switch
again between internal and
external outlet. If fresh gas
flow is not available use
emergency breathing bag.
Call DrägerService.
Warning 044 Fresh-gas flow
<3L/min
DIVA protection system
failed and fresh-gas flow
lower than 3 L/min.
Set fresh gas flow higher
than 3L/min. NOTE:At
lower flows DIVA will be ter-
minated.
Note 005 Gas supply sens.
failure
Sensor for cylinder pressure
failed.
Check sensor plug. Ex-
change high pressure sen-
sor. Call DrägerService.
Note 050 GP1 check zero Pressure transducer for
specified parameter requires
zeroing. The blood pressure
zero of monitor may not cor-
respond to the peripheral de-
vice.
Zero the transducer.
Caution 090 GP1 D high Invasive pressure value out-
side alarm limits.
Check patient and take ap-
propriate action. Change
the alarm limits if required.
Check equipment and re-
place if necessary.
Caution 090 GP1 D low Invasive pressure value out-
side alarm limits.
Check patient and take ap-
propriate action. Change
the alarm limits if required.
Check equipment and re-
place if necessary.
Classifi-
cation
Pri-
ority
Alarm text Cause Remedy