Fault – Cause – Remedy
274
Instructions for use Zeus Infinity Empowered SW 2.n
Note 025 GP1 hardware fail Invasive blood pressure
hardware failure.
Check hardware and re-
place. Call DrägerService.
Caution 090 GP1 M high Invasive pressure value out-
side alarm limits.
Check patient and take ap-
propriate action. Change
the alarm limits if required.
Check equipment and re-
place if necessary.
Caution 100 GP1 M low Invasive pressure value out-
side alarm limits.
Check patient and take ap-
propriate action. Change
the alarm limits if required.
Check equipment and re-
place if necessary.
Note 150 GP1 nonstatic Press. transducer zeroing
failed because of: Excessive
signal noise A non-static
waveform
Keep all tubing motionless,
then rezero. Change the
transducer. Check 3-way
tap, then rezero.
Caution 050 GP1 out of range Invasive pressure signal out
of measurement range.
Check patient and take ap-
propriate action. Access the
Pressure Labels menu and
assign correct label. Check
equipment and replace if
necessary.
Caution 125 GP1 S high Invasive pressure value out-
side alarm limits.
Check patient and take ap-
propriate action. Change
the alarm limits if required.
Check equipment and re-
place if necessary.
Caution 130 GP1 S low Invasive pressure value out-
side alarm limits.
Check patient and take ap-
propriate action. Change
the alarm limits if required.
Check equipment and re-
place if necessary.
Classifi-
cation
Pri-
ority
Alarm text Cause Remedy