Instructions for use Zeus Infinity Empowered SW 2.n 275
Fault – Cause – Remedy
Warning 250 GP1 static Static pressure: -No pulse,
-Press. transducer turned
off, -Catheter tip lodged
against vessel wall, -Clot on
the catheter tip
Check patient and take ap-
propriate action. Open sys-
tem by turning the 3-way
tap. Follow procedure for
dislodging catheters. Fol-
low procedure for clotted
catheters.
Note 050 GP1 transducer fail Pressure transducer failed.
Invasive pressure cannot be
measured.
Check transducer and con-
nections. Exchange trans-
ducer.
Note 025 GP1 transducer
unplugged
Pressure transducer for
specified parameter is either
unplugged or defective.
Active pressure: reconnect
or replace the cable. Inac-
tive pressure: turn off
alarms.
Note 050 GP1 zeroing Mercury calibration in
progress.
Complete calibration before
you begin monitoring the
patient.
Note 050 GP2 check zero Pressure transducer for
specified parameter requires
zeroing. The blood pressure
zero of monitor may not cor-
respond to the peripheral de-
vice.
Zero the transducer.
Caution 090 GP2 D high Invasive pressure value out-
side alarm limits.
Check patient and take ap-
propriate action. Change
the alarm limits if required.
Check equipment and re-
place if necessary.
Caution 090 GP2 D low Invasive pressure value out-
side alarm limits.
Check patient and take ap-
propriate action. Change
the alarm limits if required.
Check equipment and re-
place if necessary.