Instructions for use Zeus Infinity Empowered SW 2.n 277
Fault – Cause – Remedy
Warning 250 GP2 static Static pressure: -No pulse,
-Press. transducer turned
off, -Catheter tip lodged
against vessel wall, -Clot on
the catheter tip
Check patient and take ap-
propriate action. Open sys-
tem by turning the 3-way
tap. Follow procedure for
dislodging catheters. Fol-
low procedure for clotted
catheters.
Note 050 GP2 transducer fail Pressure transducer failed.
Invasive pressure cannot be
measured.
Check transducer and con-
nections. Exchange trans-
ducer.
Note 025 GP2 transducer
unplugged
Pressure transducer for
specified parameter is either
unplugged or defective.
Active pressure: reconnect
or replace the cable. Inac-
tive pressure: turn off
alarms.
Note 050 GP2 zeroing Mercury calibration in
progress.
Complete calibration before
you begin monitoring the
patient.
Caution 125 Heart rate/Pulse
high
Heart rate or pulse rate have
exceeded the upper alarm
limit or measured values are
above measurement range.
Observe the patient. Treat
the patient, if necessary.
Adjust alarm limit and, if
necessary, change pulse
source.
Warning 255 Heart rate/Pulse
low
Heart rate or pulse rate have
exceeded the lower alarm
limit.
Observe the patient. Treat
the patient, if necessary.
Adjust alarm limit.
Note 100 HemoMed pod 1
disconnected
Hemodynamic module at
connector 'HemoMed' not
connected during IBP, C.O.
or TEMP. measurement.
Check and change He-
moMed pod.
Call DrägerService.
Classifi-
cation
Pri-
ority
Alarm text Cause Remedy