Fault – Cause – Remedy
278
Instructions for use Zeus Infinity Empowered SW 2.n
Note 105 HemoMed pod 1
HW fail
Hemodynamic module at
connector 'HemoMed' failed
during IBP, C.O. or TEMP.
measurement.
Check and change module.
Call DrägerService.
Note 100 HemoMed pod 2
disconnected
Hemodynamic module at
connector 'Aux/Hemo 2' not
connected during IBP, C.O.
or TEMP. measurement.
Check and change cables.
Call DrägerService.
Note 105 HemoMed pod 2
HW fail
Hemodynamic module at
connector 'Aux/Hemo 2'
failed during IBP, C.O. or
TEMP. measurement.
Check and change module.
Call DrägerService.
Note 100 HemoMed pod 3
disconnected
Hemodynamic module at
connector 'Aux/Hemo 3' not
connected during BP, C.O.
or TEMP. measurement.
Check and change cables.
Call DrägerService.
Note 105 HemoMed pod 3
HW fail
Hemodynamic module at
connector 'Aux/Hemo 3'
failed during BP, C.O. or
TEMP. measurement.
Check and change module.
Call DrägerService.
Note 050 ICP check zero Pressure transducer for
specified parameter requires
zeroing. The blood pressure
zero of monitor may not cor-
respond to the peripheral de-
vice.
Zero the transducer.
Note 025 ICP hardware fail Invasive blood pressure
hardware failure.
Check hardware and re-
place. Call DrägerService.
Caution 100 ICP M high Invasive pressure value out-
side alarm limits.
Check patient and take ap-
propriate action. Change
the alarm limits if required.
Check equipment and re-
place if necessary.
Classifi-
cation
Pri-
ority
Alarm text Cause Remedy