Instructions for use Zeus Infinity Empowered SW 2.n 279
Fault – Cause – Remedy
Caution 090 ICP M low Invasive pressure value out-
side alarm limits.
Check patient and take ap-
propriate action. Change
the alarm limits if required.
Check equipment and re-
place if necessary.
Note 150 ICP nonstatic Press. transducer zeroing
failed because of: Excessive
signal noise A non-static
waveform
Keep all tubing motionless,
then rezero. Change the
transducer. Check 3-way
tap, then rezero.
Caution 050 ICP out of range Invasive pressure signal out
of measurement range.
Check patient and take ap-
propriate action. Access the
Pressure Labels menu and
assign correct label. Check
equipment and replace if
necessary.
Note 050 ICP transducer fail Pressure transducer failed.
Invasive pressure cannot be
measured.
Check transducer and con-
nections. Exchange trans-
ducer.
Note 025 ICP transducer un-
plugged
Pressure transducer for
specified parameter is either
unplugged or defective.
Active pressure: reconnect
or replace the cable. Inac-
tive pressure: turn off
alarms.
Note 050 ICP zeroing Mercury calibration in
progress.
Complete calibration before
you begin monitoring the
patient.
Note 250 Infusion 1 rate high Pump 1 Infusion Rate Limit
exceeded
Check the infusion rate of
the pump. Adjust settings if
necessary.
Note 040 Infusion 1 stopped The infusion has been
stopped or a stopped pump
has been connected.
Start infusion, check pump
settings.
Note 250 Infusion 2 rate high Pump 2 Infusion Rate Limit
exceeded
Check the infusion rate of
the pump. Adjust settings if
necessary.
Classifi-
cation
Pri-
ority
Alarm text Cause Remedy