Fault – Cause – Remedy
280
Instructions for use Zeus Infinity Empowered SW 2.n
Note 040 Infusion 2 stopped The infusion has been
stopped or a stopped pump
has been connected.
Start infusion, check pump
settings.
Note 250 Infusion 3 rate high Pump 3 Infusion Rate Limit
exceeded
Check the infusion rate of
the pump. Adjust settings if
necessary.
Note 040 Infusion 3 stopped The infusion has been
stopped or a stopped pump
has been connected.
Start infusion, check pump
settings.
Note 250 Infusion 4 rate high Pump 4 Infusion Rate Limit
exceeded
Check the infusion rate of
the pump. Adjust settings if
necessary.
Note 040 Infusion 4 stopped The infusion has been
stopped or a stopped pump
has been connected.
Start infusion, check pump
settings.
Warning 100 Infusion line 1
blocked
Occlusion at line 1 detected.
Pressure limit reached. Infu-
sion 1 stopped.
Check infusion line 1 and
connection to the patient.
Increase pressure alarm
limit. Restart infusion.
Caution 100 Infusion line 1
blocked?
The line 1 pressure in-
creased rapidly. Pressure
alarm limit is nearly reached.
Check infusion line and
connection to the patient.
Increase pressure alarm
limit, if necessary.
Warning 100 Infusion line 2
blocked
Occlusion at line 2 detected.
Pressure limit reached. Infu-
sion 2 stopped.
Check infusion line 2 and
connection to the patient.
Increase pressure alarm
limit. Restart infusion 2.
Caution 100 Infusion line 2
blocked?
The line 2 pressure in-
creased rapidly. Pressure
alarm limit is nearly reached.
Check infusion line and
connection to the patient.
Increase pressure alarm
limit, if necessary.
Classifi-
cation
Pri-
ority
Alarm text Cause Remedy