Fault – Cause – Remedy
290
Instructions for use Zeus Infinity Empowered SW 2.n
Note 050 LV zeroing Mercury calibration in
progress.
Complete calibration before
you begin monitoring the
patient.
Caution 000 Mains supply fail-
ure
Power failure. A fully
charged battery allows at
least 30 min of operation.
Restore central power sup-
ply. Check circuit breakers.
Check battery capacity dis-
play. If capacity is < 10 %
the device may switch off!
Caution 020 Minute volume high Upper alarm limit of minute
volume exceeded.
Check alarm limit. Check
minute volume and respira-
tory rate.
Caution 200 Minute volume low Exp. minute volume lower
than lower limit. Leakage in
breathing system; hose
blocked. Volume deficiency
during press. limitation.
Fresh-gas flow too low.
Check limit setting. Check
tube, hoses and breath.
system. Check P
max and
P
insp setting. Increase fresh
gas flow.
Warning 255 Mixer+DIVA failure Mixer and DIVA not opera-
ble. Setting of ventilator not
possible!
Open safety O
2 valve. Con-
tinue case with IV drug.
Check ventilation. Switch
'AN-VENTILATOR OFF'.
Warning 255 Mixer+DIVA failure Mixer and DIVA not opera-
ble. Setting of ventilator not
possible!
Open safety O
2 valve. Con-
tinue case with IV drug.
Check ventilation. Switch
'AN-VENTILATOR OFF'.
Note 100 MultiMed Pod dis-
connected
MultiMed cable not con-
nected or faulty during 3, 5,
or 6-lead monitoring.
Check cables and connec-
tions and replace if neces-
sary.
Note 105 MultiMed Pod HW
failure
MultiMed cable failed during
3, 5, or 6-lead monitoring.
Check and change
MultiMed cable. Call
DrägerService.
Note 050 N
2O central supply
high
N2O supply pressure too
high. N2O-pressure
>8x 100 kPa.
Check central supply.
Classifi-
cation
Pri-
ority
Alarm text Cause Remedy