EasyManua.ls Logo

Dräger Zeus Infinity Empowered - Page 289

Dräger Zeus Infinity Empowered
453 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instructions for use Zeus Infinity Empowered SW 2.n 289
Fault – Cause – Remedy
Note 150 LV nonstatic Press. transducer zeroing
failed because of: Excessive
signal noise A non-static
waveform
Keep all tubing motionless,
then rezero. Change the
transducer. Check 3-way
tap, then rezero.
Caution 050 LV out of range Invasive pressure signal out
of measurement range.
Check patient and take ap-
propriate action. Access the
Pressure Labels menu and
assign correct label. Check
equipment and replace if
necessary.
Caution 125 LV S high Invasive pressure value out-
side alarm limits.
Check patient and take ap-
propriate action. Change
the alarm limits if required.
Check equipment and re-
place if necessary.
Caution 130 LV S low Invasive pressure value out-
side alarm limits.
Check patient and take ap-
propriate action. Change
the alarm limits if required.
Check equipment and re-
place if necessary.
Warning 250 LV static Static pressure: -No pulse,
-Press. transducer turned
off, -Catheter tip lodged
against vessel wall, -Clot on
the catheter tip
Check patient and take ap-
propriate action. Open sys-
tem by turning the 3-way
tap. Follow procedure for
dislodging catheters. Fol-
low procedure for clotted
catheters.
Note 050 LV transducer fail Pressure transducer failed.
Invasive pressure cannot be
measured.
Check transducer and con-
nections. Exchange trans-
ducer.
Note 025 LV transducer un-
plugged
Pressure transducer for
specified parameter is either
unplugged or defective.
Active pressure: reconnect
or replace the cable. Inac-
tive pressure: turn off
alarms.
Classifi-
cation
Pri-
ority
Alarm text Cause Remedy

Table of Contents

Related product manuals