Fault – Cause – Remedy
288
Instructions for use Zeus Infinity Empowered SW 2.n
Note 020 LL electrode fail Lead-off condition for indi-
cated lead detected: -Broken
cable – Loose lead wire
-Faulty lead wire – Wrong
lead -Dried out gel -Bad con-
tact.
Check defective cables and
electrodes. Reapply gel to
electrodes. Make sure the
skin is properly prepped.
Select another ECG/ST
lead. Set correct type of ca-
ble.
Note 050 LV check zero Pressure transducer for
specified parameter requires
zeroing. The blood pressure
zero of monitor may not cor-
respond to the peripheral de-
vice.
Zero the transducer.
Caution 090 LV D high Invasive pressure value out-
side alarm limits.
Check patient and take ap-
propriate action. Change
the alarm limits if required.
Check equipment and re-
place if necessary.
Caution 090 LV D low Invasive pressure value out-
side alarm limits.
Check patient and take ap-
propriate action. Change
the alarm limits if required.
Check equipment and re-
place if necessary.
Note 025 LV hardware fail Invasive blood pressure
hardware failure.
Check hardware and re-
place. Call DrägerService.
Caution 090 LV M high Invasive pressure value out-
side alarm limits.
Check patient and take ap-
propriate action. Change
the alarm limits if required.
Check equipment and re-
place if necessary.
Caution 100 LV M low Invasive pressure value out-
side alarm limits.
Check patient and take ap-
propriate action. Change
the alarm limits if required.
Check equipment and re-
place if necessary.
Classifi-
cation
Pri-
ority
Alarm text Cause Remedy