Instructions for use Zeus Infinity Empowered SW 2.n 287
Fault – Cause – Remedy
Caution 050 LA out of range Invasive pressure signal out
of measurement range.
Check patient and take ap-
propriate action. Access the
Pressure Labels menu and
assign correct label. Check
equipment and replace if
necessary.
Note 050 LA transducer fail Pressure transducer failed.
Invasive pressure cannot be
measured.
Check transducer and con-
nections. Exchange trans-
ducer.
Note 025 LA transducer un-
plugged
Pressure transducer for
specified parameter is either
unplugged or defective.
Active pressure: reconnect
or replace the cable. Inac-
tive pressure: turn off
alarms.
Note 050 LA zeroing Mercury calibration in
progress.
Complete calibration before
you begin monitoring the
patient.
Note 000 Leads invalid Lead-off condition detected:
-Broken cable - Loose lead
wire -Faulty lead wire
- Wrong lead -Dried out gel
-Bad contact.
Check defective cables and
electrodes. Reapply gel to
electrodes. Make sure the
skin is properly prepped.
Select another ECG/ST
lead. Set correct type of ca-
ble.
Caution 045 Leak or fresh gas
low
Fresh gas deficiency!. Tidal
volume cannot not be deliv-
ered. Leak in breathing cir-
cle.
Increase fresh gas flow.
Check O
2- and vol. agent
settings. Check breathing
circle and manual breathing
bag for leakage.
Note 100 Leakage or fresh
gas low
Fresh gas deficiency! Man-
ual breathing bag not filled
sufficiently. Leak in breathing
circle.
Increase fresh gas flow.
Check O
2- and vol. agent
settings. Check breathing
circle and manual breathing
bag for leakage.
Classifi-
cation
Pri-
ority
Alarm text Cause Remedy