Fault – Cause – Remedy
286
Instructions for use Zeus Infinity Empowered SW 2.n
Note 100 KVO pump 4 Volume limit of infusion 4 has
been reached. 'Keep vein
open' function is activated.
Infusion continues with a low
infusion rate (1 mL/h).
Check infusion. Check vol-
ume alarm limit. To change
KVO configuration at the
pump refer to the module
DPS instructions.
Note 050 LA check zero Pressure transducer for
specified parameter requires
zeroing. The blood pressure
zero of monitor may not cor-
respond to the peripheral de-
vice.
Zero the transducer.
Note 020 LA electrode fail Lead-off condition for indi-
cated lead detected: -Broken
cable - Loose lead wire
- Faulty lead wire - Wrong
lead -Dried out gel -Bad con-
tact.
Check defective cables and
electrodes. Reapply gel to
electrodes. Make sure the
skin is properly prepped.
Select another ECG/ST
lead. Set correct type of ca-
ble.
Note 025 LA hardware fail Invasive blood pressure
hardware failure.
Check hardware and re-
place. Call DrägerService.
Caution 090 LA M high Invasive pressure value out-
side alarm limits.
Check patient and take ap-
propriate action. Change
the alarm limits if required.
Check equipment and re-
place if necessary.
Caution 100 LA M low Invasive pressure value out-
side alarm limits.
Check patient and take ap-
propriate action. Change
the alarm limits if required.
Check equipment and re-
place if necessary.
Note 150 LA nonstatic Press. transducer zeroing
failed because of: Excessive
signal noise A non-static
waveform
Keep all tubing motionless,
then rezero. Change the
transducer. Check 3-way
tap, then rezero.
Classifi-
cation
Pri-
ority
Alarm text Cause Remedy