Instructions for use Zeus Infinity Empowered SW 2.n 285
Fault – Cause – Remedy
Caution 255 Insp. vol. agent
high
Insp. vol. agent mixture
higher than 3 xMAC.
Check DIVA settings.
Check fresh-gas flow.
Caution 250 Internal safety fan
failed
The safety fan for internal
gas evacuation failed. If
there is an internal leakage,
the O
2 concentration in the
interior of the device may in-
crease. Danger of fire!
Shut down as soon as pos-
sible. Call DrägerService.
Warning 100 IV remote control
failed
Remote control of infusion
was terminated by the safety
system! Infusion is contin-
ued.
Check infusion setting at
pump. Control infusion at
pump. To restart remote
control System needs to be
restarted.
Caution 200 IV system discon-
nected
Communication error to
IV-System. Cable discon-
nected.
Check connection of the
IVDock to system, replace
cable or IVDock. Call
DrägerService.
Note 125 IVDock incompati-
ble
IVDock has unknown ver-
sion.
Exchange IVDock. Control
infusion directly at the
pump. Call DrägerService.
Note 100 KVO pump 1 Volume limit of infusion 1 has
been reached. 'Keep vein
open' function is activated.
Infusion continues with a low
infusion rate (1 mL/h).
Check infusion. Check vol-
ume alarm limit. To change
KVO configuration at the
pump refer to the module
DPS instructions.
Note 100 KVO pump 2 Volume limit of infusion 2 has
been reached. 'Keep vein
open' function is activated.
Infusion continues with a low
infusion rate (1 mL/h).
Check infusion. Check vol-
ume alarm limit. To change
KVO configuration at the
pump refer to the module
DPS instructions.
Note 100 KVO pump 3 Volume limit of infusion 3 has
been reached. 'Keep vein
open' function is activated.
Infusion continues with a low
infusion rate (1 mL/h).
Check infusion. Check vol-
ume alarm limit. To change
KVO configuration at the
pump refer to the module
DPS instructions.
Classifi-
cation
Pri-
ority
Alarm text Cause Remedy