Fault – Cause – Remedy
284
Instructions for use Zeus Infinity Empowered SW 2.n
Caution 000 Insp. oxygen high Upper limit of insp. Oxygen-
concentration exceeded for
2 breath. cycles.
Check Oxygen delivery set-
ting! Check fresh gas flow!
Warning 180 Insp. oxygen low Inspiratory Oxygen concen-
tration fell below lower limit
for 2 breath. cycles.
Check O
2 supply! Check
Oxygen delivery setting!
Check fresh gas flow!
Check SpO
2 concentration!
Note 175 Insp. pressure sen-
sor fail
Failure of insp. pressure
sensor. Measurement of
breathing pressure may be
inaccurate. Breathing press.
is measured by exp. sensor.
Check breathing pressure!
Call DrägerService.
Caution 160 Insp. resistance
high
Inspiratory resistance is in-
creased compared to self-
test – Stenosis or high resist-
ance in insp. hose, tube or fil-
ter.
Check insp. hose, tube and
filter.
Caution 255 Insp. sevoflurane
high
Insp. vol. agent higher than
alarm limit for 30 sec.
Check volatile agent deliv-
ery setting! Check fresh gas
flow!
Caution 255 Insp. sevoflurane
high
Insp. vol. agent mixture
higher than 3 xMAC.
Check DIVA settings and
fresh gas flow.
Note 255 Insp. sevoflurane
high
Insp. vol. agent higher than
alarm limit for 30 sec.
Check volatile agent deliv-
ery setting! Check fresh gas
flow!
Caution 060 Insp. sevoflurane
low
Inspiratory Sevoflurane con-
centration fell below lower
limit for 2 breath. cycles.
Check volatile agent deliv-
ery setting! Check fresh gas
flow! Check function and
liquid level of DIVA. Check
CO
2-sample line for leak-
age!
Caution 254 Insp. vol. agent
high
Volatile agent in inspiration
higher than setpoint.
Use Fresh-gas control
mode.
Classifi-
cation
Pri-
ority
Alarm text Cause Remedy