Instructions for use Zeus Infinity Empowered SW 2.n 283
Fault – Cause – Remedy
Caution 060 Insp. halothane low Inspiratory Halothane con-
centration fell below lower
limit for 2 breath. cycles.
Check volatile agent deliv-
ery setting! Check fresh gas
flow! Check function and
liquid level of DIVA. Check
CO
2-sample line for leak-
age!
Caution 255 Insp. isoflurane
high
Insp. vol. agent higher than
alarm limit for 30 sec.
Check volatile agent deliv-
ery setting! Check fresh gas
flow!
Caution 255 Insp. isoflurane
high
Insp. vol. agent mixture
higher than 3 xMAC.
Check DIVA settings and
fresh gas flow.
Note 255 Insp. isoflurane
high
Insp. vol. agent higher than
alarm limit for 30 sec.
Check volatile agent deliv-
ery setting! Check fresh gas
flow!
Caution 060 Insp. isoflurane low Inspiratory Isoflurane con-
centration fell below lower
limit for 2 breath. cycles.
Check volatile agent deliv-
ery setting! Check fresh gas
flow! Check function and
liquid level of DIVA. Check
CO
2-sample line for leak-
age!
Caution 000 Insp. MAC > 3 Insp. vol. agent concentra-
tion too high. Insp.3 MAC ex-
ceeded.
Check DIVA settings and
fresh gas flow.
Caution 255 Insp. MAC > 5 Insp. vol. agent concentra-
tion too high. Insp.5 MAC ex-
ceeded.
Check DIVA settings and
fresh gas flow.
Caution 020 Insp. N
2O high Upper limit of insp. N2O-
concentration exceeded for
2 breath. cycles.
Check O2-setting! Check
fresh gas flow! Flush with
high fresh gas flow!
Caution 000 Insp. N
2O low Inspiratory N2O concentra-
tion fell below lower limit for
2 breath. cycles.
Check N2O supply! Check
O
2-setting! Check fresh gas
flow!
Warning 180 Insp. O
2 low Protect. System: Insp. O2
concentration lower than
21 %!.
Check O2 concentration.
Check O2 supply. Use
Fresh-gas control mode.
Classifi-
cation
Pri-
ority
Alarm text Cause Remedy