Fault – Cause – Remedy
282
Instructions for use Zeus Infinity Empowered SW 2.n
Note 255 Insp. desflurane
high
Insp. vol. agent higher than
alarm limit for 30 sec.
Check volatile agent deliv-
ery setting! Check fresh gas
flow!
Caution 060 Insp. desflurane
low
Inspiratory Desflurane con-
centration fell below lower
limit for 2 breaths.
Check volatile agent deliv-
ery setting! Check fresh gas
flow! Check function and
liquid level of DIVA. Check
CO
2-sample line for leak-
age!
Caution 255 Insp. enflurane
high
Insp. vol. agent higher than
alarm limit for 30 sec.
Check DIVA settings and
fresh gas flow.
Caution 255 Insp. enflurane
high
Insp. vol. agent mixture
higher than 3 xMAC.
Check DIVA settings and
fresh gas flow.
Note 255 Insp. enflurane
high
Insp. vol. agent higher than
alarm limit for 30 sec.
Check DIVA settings and
fresh gas flow.
Caution 060 Insp. enflurane low Inspiratory Enflurane con-
centration fell below lower
limit for 2 breath. Cycles.
Check volatile agent deliv-
ery setting! Check fresh gas
flow! Check function and
liquid level of DIVA. Check
CO
2-sample line for leak-
age!
Note 180 Insp. flow sensor
failure
Failure of inspiratory flow
sensor: – Sensor not con-
nected or defective
Perform leaktest incl. au-
tom. calibration! Exchange
sensor and perform
leaktest!
Caution 255 Insp. halothane
high
Insp. vol. agent higher than
alarm limit for 30 sec.
Check volatile agent deliv-
ery setting! Check fresh gas
flow!
Caution 255 Insp. halothane
high
Insp. vol. agent mixture
higher than 3 xMAC.
Check DIVA settings and
fresh gas flow.
Note 255 Insp. halothane
high
Insp. vol. agent higher than
alarm limit for 30 sec.
Check volatile agent deliv-
ery setting! Check fresh gas
flow!
Classifi-
cation
Pri-
ority
Alarm text Cause Remedy