Fault – Cause – Remedy
292
Instructions for use Zeus Infinity Empowered SW 2.n
Caution 090 NIBP D high Non invasive pressure value
outside alarm limits.
Check patient and take ap-
propriate action. Change
the alarm limits if required.
Check equipment and re-
place if necessary.
Caution 090 NIBP D low Non invasive pressure value
outside alarm limits.
Check patient and take ap-
propriate action. Change
the alarm limits if required.
Check equipment and re-
place if necessary.
Note 050 NIBP hardware
fault
Non invasive blood pressure
measurement failed.
Call DrägerService
Note 050 NIBP hose un-
plugged
NIBP hose is unplugged. Reconnect the hose.
Caution 200 NIBP inflation fail Systolic pressure is higher
than inflation limit. Pump or
valves have been energized
for longer than 2 minutes
(90 sec. neonatal mode).
Select higher inflation limit
setting. Check hose and
cuff for obstructions. Dis-
connect and reconnect cuff.
Replace if necessary. Call
DrägerService.
Caution 090 NIBP M high Non invasive pressure value
outside alarm limits.
Check patient and take ap-
propriate action. Change
the alarm limits if required.
Check equipment and re-
place if necessary.
Caution 100 NIBP M low Non invasive pressure value
outside alarm limits.
Check patient and take ap-
propriate action. Change
the alarm limits if required.
Check equipment and re-
place if necessary.
Caution 200 NIBP open line Inflation time during cuff in-
flation cycle is too long or in-
flation rate is too low.
Check to ensure that the
hose and cuff are properly
connected to the monitor.
Check the cuff size to be
suitable for the patient.
Classifi-
cation
Pri-
ority
Alarm text Cause Remedy