Instructions for use Zeus Infinity Empowered SW 2.n 293
Fault – Cause – Remedy
Caution 250 NIBP press exceed Cuff overpressure circuit has
failed.
Do not use NIBP measure-
ment. Call DrägerService.
Caution 210 NIBP press leak NIBP hose/cuff hose discon-
nected? Leakage in NIBP
hose/cuff?
Check connections! Re-
place cuff or hose!
Caution 125 NIBP S high Non invasive pressure value
outside alarm limits.
Check patient and take ap-
propriate action. Change
the alarm limits if required.
Check equipment and re-
place if necessary.
Caution 130 NIBP S low Non invasive pressure value
outside alarm limits.
Check patient and take ap-
propriate action. Change
the alarm limits if required.
Check equipment and re-
place if necessary.
Caution 200 NIBP time exceed A measurement took longer
than 2 minutes (adult or
pediatric) or 90 seconds (ne-
onatal) and was aborted
(e.g. motion artefact).
Repeat the measurement.
Note 100 NIBP zero out of
range
Monitor is unable to zero
transducer within 30 sec-
onds from start of NIBP pro-
gram, usually because of
motion artefact.
Check the patient and take
appropriate action. Move
cuff to a limb with less
movement. Call
DrägerService.
Note 100 NMT artefact Excessive signal noise due
to patient´s move-
ment/tremor or interference
from auxiliary equipment.
Check electrodes. Reapply,
if necessary. Isolate patient
from auxiliary equipment, if
necessary. Isolate patient
from extraneous movement
or vibration.
Note 250 NMT check elec-
trodes
Current cannot be delivered.
Electrodes or connections
not properly placed/function-
ing.
Check electrode. Reapply,
if necessary. Check cable
connections.
Classifi-
cation
Pri-
ority
Alarm text Cause Remedy