Fault – Cause – Remedy
294
Instructions for use Zeus Infinity Empowered SW 2.n
Note 200 NMT check sensor Sensor malfunctioning or
disconnected.
Check sensor application.
Replace accelerometer
sensor, if necessary.
Note 100 NMT check temp
sensor
Temperature cable broken or
missing.
Check temperature sensor.
Replace, if necessary.
Note 125 NMT initializing Pod is performing power-up
tests. Supramaximal current
and reference twitch being
detected. Pod is re-calibrat-
ing due to a rapid change in
twitch amplitude.
Wait until self tests are com-
pleted. Wait until references
are established.
Note 150 NMT single out of
range
Single mode parameter
value falls outside the meas-
uring range (high).
Check patient. Treat, if nec-
essary.
Note 150 NMT temp out of
range
Temperature value falls out-
side the measuring range
(high/low).
Check patient. Treat, if nec-
essary. Check sensor. Re-
place, if necessary.
Note 150 NMT temp out of
range
Temperature value falls out-
side the measuring range
(high/low).
Check patient. Treat, if nec-
essary. Check sensor. Re-
place, if necessary.
Note 100 NMT using internal
reference
Reference twitch could not
be established. Patient might
have been relaxed before
reference was established.
Check electrodes and sen-
sor. Continue measurement
using internal reference or
change to TOF mode.
Note 250 O
2 central supply
low
O2 central supply pressure is
low. Wall outlet is faulty. O2
plug is not connected to wall
outlet.
Check if O2 is delivered to
the patient. Open O2 cylin-
der. Remove O
2 plug from
wall outlet. Connect O2 plug
to wall outlet.
Warning 100 O
2 delivery low ac-
curacy
O2 delivery is working, but
not within the specified toler-
ance. A large deviation may
lead to the termination of
N
2O or Air delivery.
Check inspiratory O
2
concentration. Call
DrägerService.
Note 250 O2 sensor failure O2-Sensor failed. Auto con-
trol delivery is not available.
Ensure sufficient oxygena-
tion. Call DrägerService.
Note 050 Oxygen central
supply high
Oxygen supply pressure too
high. O
2-pressure
>8x 100 kPa.
Check central supply.
Classifi-
cation
Pri-
ority
Alarm text Cause Remedy