Instructions for use Zeus Infinity Empowered SW 2.n 295
Fault – Cause – Remedy
Warning 134 Oxygen cylinder
empty
O2 cylinder empty or not
open (<20*100 kPa) and O2
supply lower than
1.8*100 kPa! Mixer contin-
ues with Air delivery!
Reduce fresh-gas flow, use
Auto control mode. Open or
exchange O
2 cylinder.
Check insp. O2 concentra-
tion. Check O2 supply.
Note 255 Oxygen cylinder
low
O
2-cylinder pressure low. O2
cylinder nearly emptied.
Use filled O2 cylinder. Re-
store central supply.
Note 005 Oxygen cylinder
sensor fail
O
2 high pressure sensor not
connected. Sensor failure.
Connect sensor. Check
sensor. Call DrägerService.
Warning 135 Oxygen supply fail Oxygen supply failed. Hose
not plugged into wall socket
or hose kinked.
Open O
2-cylinder. Check
central supply. Check wall
connector.
Note 254 Oxygen supply low Oxygen supply too low. O
2
supply between 180 and
260 kPa. O2-delivery not
within specification.
Check O2 supply. Prepare
O2 backup cylinders.
Note 050 PA check zero Pressure transducer for
specified parameter requires
zeroing. The blood pressure
zero of monitor may not cor-
respond to the peripheral de-
vice.
Zero the transducer.
Caution 090 PA D high Invasive pressure value out-
side alarm limits.
Check patient and take ap-
propriate action. Change
the alarm limits if required.
Check equipment and re-
place if necessary.
Caution 090 PA D low Invasive pressure value out-
side alarm limits.
Check patient and take ap-
propriate action. Change
the alarm limits if required.
Check equipment and re-
place if necessary.
Note 025 PA hardware fail Invasive blood pressure
hardware failure.
Check hardware and re-
place. Call DrägerService.
Caution 090 PA M high Invasive pressure value out-
side alarm limits.
Check patient and take ap-
propriate action. Change
the alarm limits if required.
Check equipment and re-
place if necessary.
Classifi-
cation
Pri-
ority
Alarm text Cause Remedy