Fault – Cause – Remedy
298
Instructions for use Zeus Infinity Empowered SW 2.n
Caution 250 Power supply temp
high
Power supply temperature
too high.
Minimize power consump-
tion. Ventilate manually.
Switch off device as soon
as possible.
Call DrägerService.
Caution 000 Press.supp. time
limited
Pressure support cycle was
aborted. The OFF criterion
(flow <25 % of max. insp.
flow) was not reached within
4 sec.
Check for tube/mask leak-
age. Check patient settings.
Use Pressure Control.
Note 000 Printer error Could not access any printer. Choose another printer, if
available.
Call DrägerService.
Caution 100 Protect sensor war-
mup
Protection sensor is not
available. Agent delivery re-
quires a fresh gas flow
>3 L/min.
Latest after 1/2 hour the
protect sensor calibrates
again. After this the delivery
may be available again.
Use fresh gas flow
>3 L/min.
Note 255 Protect water trap
blocked
Protect water trap or gas
measurement system
blocked. Protect water trap
full or expired.
Check Protect water trap
and gas measurement sys-
tem – exchange, if neces-
sary. Exchange water trap
at least every 4 weeks.
Caution 010 Protect water trap? Protect water trap or gas
measurement system
blocked. Protect water trap
full or expired.
Check Protect water trap
and gas measurement sys-
tem – exchange, if neces-
sary. Exchange water trap
at least every 4 weeks.
Caution 040 Pump 1 battery
empty
Infusion 1 stopped. Battery
of pump 1 discharged.
Reconnect IVDock to sys-
tem for power supply or ex-
change battery. Call
DrägerService.
Note 040 Pump 1 battery low Battery nearly empty. Reconnect IVDock to sys-
tem for power supply or ex-
change battery. Call
DrägerService.
Caution 200 Pump 1 failure Infusion 1 stopped. Pump
detected internal failure.
Exchange pump 1. Call
DrägerService.
Note 120 Pump 1 incompati-
ble
Pump 1 hardware or soft-
ware version not compatible.
Remote control not possible.
Control infusion at pump.
Replace pump for remote
control.
Classifi-
cation
Pri-
ority
Alarm text Cause Remedy