Instructions for use Zeus Infinity Empowered SW 2.n 299
Fault – Cause – Remedy
Note 080 Pump 1 infusion
limit
Infusion limit is nearly
reached for pump 1. If limit is
reached infusion will be
stopped!
Prepare next syringe if nec-
essary. Increase alarm limit
if needed.
Caution 150 Pump 1 infusion
limit
Upper alarm limit for infusion
volume 1 exceeded. Infusion
stopped.
Check syringe 1. Check
alarm limits and increase
limit if necessary.
Caution 200 Pump 1 operation
failed
Remote control, setting or
start/stop of pump 1 failed.
Check last setting of pump
1. Repeat setting or start of
pump, if necessary.
Caution 060 Pump 1 pressure
drop
Pressure drop in syringe 1
detected. Infusion line is
leaking.
Check infusion line and
connections at patient and
syringe side. Be aware of
leakages. Deactivate the
DPS- function at the pump.
Caution 060 Pump 1 pressure
rise
Pressure rise in syringe 1
detected. Infusion line
kinked or blocked.
Check infusion line 1 and
syringe 1. Deactivate the
DPS- function at the pump,
if appropriate.
Note 125 Pump 1 removed Pump 1 removed during run-
ning infusion.
Control infusion at the
pump. To restart remote
control reconnect pump.
Note 125 Pump 1 removed Pump 1 removed in stopped
mode.
To restart remote control re-
connect pump.
Note 100 Pump 1 unlocked Pump 1 is unlocked. Lock the pump.
Caution 040 Pump 2 battery
empty
Infusion 2 stopped. Battery
of pump 2 discharged.
Reconnect IVDock to sys-
tem for power supply or ex-
change battery. Call
DrägerService.
Note 040 Pump 2 battery low Battery nearly empty. Reconnect IVDock to sys-
tem for power supply or ex-
change battery. Call
DrägerService.
Caution 200 Pump 2 failure Infusion 2 stopped. Pump
detected internal failure.
Exchange pump 2. Call
DrägerService.
Note 120 Pump 2 incompati-
ble
Pump 2 hardware or soft-
ware version not compatible.
Remote control not possible.
Control infusion at pump.
Replace pump for remote
control.
Classifi-
cation
Pri-
ority
Alarm text Cause Remedy