Fault – Cause – Remedy
300
Instructions for use Zeus Infinity Empowered SW 2.n
Note 080 Pump 2 infusion
limit
Infusion limit is nearly
reached for pump 2. If limit is
reached infusion will be
stopped!
Prepare next syringe if nec-
essary. Increase alarm limit
if needed.
Caution 150 Pump 2 infusion
limit
Upper alarm limit for infusion
volume 2 exceeded. Infusion
stopped.
Check syringe 2. Check
alarm limits and increase
limit if necessary.
Caution 200 Pump 2 operation
failed
Remote control, setting or
start/stop of pump 2 failed.
Check last setting of pump
2. Repeat setting or start of
pump, if necessary.
Caution 060 Pump 2 pressure
drop
Pressure drop in syringe 2
detected. Infusion line is
leaking.
Check infusion line and
connections at patient and
syringe side. Be aware of
leakages. Deactivate the
DPS- function at the pump.
Caution 060 Pump 2 pressure
rise
Pressure rise in syringe 2
detected. Infusion line
kinked or blocked.
Check infusion line 2 and
syringe 2. Deactivate the
DPS- function at the pump,
if appropriate.
Note 125 Pump 2 removed Pump 2 removed during run-
ning infusion.
Control infusion at the
pump. To restart remote
control reconnect pump.
Note 125 Pump 2 removed Pump 2 removed in stopped
mode.
To restart remote control re-
connect pump.
Note 100 Pump 2 unlocked Pump 2 is unlocked. Lock the pump.
Caution 040 Pump 3 battery
empty
Infusion 3 stopped. Battery
of pump 3 discharged.
Reconnect IVDock to sys-
tem for power supply or ex-
change battery. Call
DrägerService.
Note 040 Pump 3 battery low Battery nearly empty. Reconnect IVDock to sys-
tem for power supply or ex-
change battery. Call
DrägerService.
Caution 200 Pump 3 failure Infusion 3 stopped. Pump
detected internal failure.
Exchange pump 3. Call
DrägerService.
Note 120 Pump 3 incompati-
ble
Pump 3 hardware or soft-
ware version not compatible.
Remote control not possible.
Control infusion at pump.
Replace pump for remote
control.
Classifi-
cation
Pri-
ority
Alarm text Cause Remedy