Instructions for use Zeus Infinity Empowered SW 2.n 301
Fault – Cause – Remedy
Note 080 Pump 3 infusion
limit
Infusion limit is nearly
reached for pump 3. If limit is
reached infusion will be
stopped!
Prepare next syringe if nec-
essary. Increase alarm limit
if needed.
Caution 150 Pump 3 infusion
limit
Upper alarm limit for infusion
volume 3 exceeded. Infusion
stopped.
Check syringe 3. Check
alarm limits and increase
limit if necessary.
Caution 200 Pump 3 operation
failed
Remote control, setting or
start/stop of pump 3 failed.
Check last setting of pump
3. Repeat setting or start of
pump, if necessary.
Caution 060 Pump 3 pressure
drop
Pressure drop in syringe 3
detected. Infusion line is
leaking.
Check infusion line and
connections at patient and
syringe side. Be aware of
leakages. Deactivate the
DPS- function at the pump.
Caution 060 Pump 3 pressure
rise
Pressure rise in syringe 3
detected. Infusion line
kinked or blocked.
Check infusion line 3 and
syringe 3. Deactivate the
DPS- function at the pump,
if appropriate.
Note 125 Pump 3 removed Pump 3 removed during run-
ning infusion.
Control infusion at the
pump. To restart remote
control reconnect pump.
Note 125 Pump 3 removed Pump 3 removed in stopped
mode.
To restart remote control re-
connect pump.
Note 100 Pump 3 unlocked Pump 3 is unlocked. Lock the pump.
Caution 040 Pump 4 battery
empty
Infusion 4 stopped. Battery
of pump 4 discharged.
Reconnect IVDock to sys-
tem for power supply or ex-
change battery. Call
DrägerService.
Note 040 Pump 4 battery low Battery nearly empty. Reconnect IVDock to sys-
tem for power supply or ex-
change battery. Call
DrägerService.
Caution 200 Pump 4 failure Infusion 4 stopped. Pump
detected internal failure.
Exchange pump 4. Call
DrägerService.
Note 120 Pump 4 incompati-
ble
Pump 4 hardware or soft-
ware version not compatible.
Remote control not possible.
Control infusion at pump.
Replace pump for remote
control.
Classifi-
cation
Pri-
ority
Alarm text Cause Remedy