Fault – Cause – Remedy
302
Instructions for use Zeus Infinity Empowered SW 2.n
Note 080 Pump 4 infusion
limit
Infusion limit is nearly
reached for pump 4. If limit is
reached infusion will be
stopped!
Prepare next syringe if nec-
essary. Increase alarm limit
if needed.
Caution 150 Pump 4 infusion
limit
Upper alarm limit for infusion
volume 4 exceeded. Infusion
stopped.
Check syringe 4. Check
alarm limits and increase
limit if necessary.
Caution 200 Pump 4 operation
failed
Remote control, setting or
start/stop of pump 4 failed.
Check last setting of pump
4. Repeat setting or start of
pump, if necessary.
Caution 060 Pump 4 pressure
drop
Pressure drop in syringe 4
detected. Infusion line is
leaking.
Check infusion line and
connections at patient and
syringe side. Be aware of
leakages. Deactivate the
DPS- function at the pump.
Caution 060 Pump 4 pressure
rise
Pressure rise in syringe 4
detected. Infusion line
kinked or blocked.
Check infusion line 4 and
syringe 4. Deactivate the
DPS- function at the pump,
if appropriate.
Note 125 Pump 4 removed Pump 4 removed during run-
ning infusion.
Control infusion at the
pump. To restart remote
control reconnect pump.
Note 125 Pump 4 removed Pump 4 removed in stopped
mode.
To restart remote control re-
connect pump.
Note 100 Pump 4 unlocked Pump 4 is unlocked. Lock the pump.
Caution 000 PVC high Premature ventricular con-
tractions: PVCs per minute
have exceeded the upper
alarm limit.
Check patient and take ap-
propriate action.
Note 050 RA check zero Pressure transducer for
specified parameter requires
zeroing. The blood pressure
zero of monitor may not cor-
respond to the peripheral de-
vice.
Zero the transducer.
Note 020 RA electrode fail Lead-off condition for indi-
cated lead detected: -Broken
cable - Loose lead wire -
Faulty lead wire - Wrong lead
-Dried out gel -Bad contact.
Check defective cables and
electrodes. Reapply gel to
electrodes. Make sure the
skin is properly prepped.
Select another ECG/ST
lead. Set correct type of ca-
ble.
Classifi-
cation
Pri-
ority
Alarm text Cause Remedy