Instructions for use Zeus Infinity Empowered SW 2.n 303
Fault – Cause – Remedy
Note 025 RA hardware fail Invasive blood pressure
hardware failure.
Check hardware and re-
place. Call DrägerService.
Caution 090 RA M high Invasive pressure value out-
side alarm limits.
Check patient and take ap-
propriate action. Change
the alarm limits if required.
Check equipment and re-
place if necessary.
Caution 100 RA M low Invasive pressure value out-
side alarm limits.
Check patient and take ap-
propriate action. Change
the alarm limits if required.
Check equipment and re-
place if necessary.
Note 150 RA nonstatic Press. transducer zeroing
failed because of: Excessive
signal noise A non-static
waveform
Keep all tubing motionless,
then rezero. Change the
transducer. Check 3-way
tap, then rezero.
Caution 050 RA out of range Invasive pressure signal out
of measurement range.
Check patient and take ap-
propriate action. Access the
Pressure Labels menu and
assign correct label. Check
equipment and replace if
necessary.
Note 050 RA transducer fail Pressure transducer failed.
Invasive pressure cannot be
measured.
Check transducer and con-
nections. Exchange trans-
ducer.
Note 025 RA transducer un-
plugged
Pressure transducer for
specified parameter is either
unplugged or defective.
Active pressure: reconnect
or replace the cable. Inac-
tive pressure: turn off
alarms.
Note 050 RA zeroing Mercury calibration in
progress.
Complete calibration before
you begin monitoring the
patient.
Note 000 Relearn Monitor has begun 10-sec-
ond capture of rest ECG
waveforms.
Prevent movement of the
patient.
Warning 030 Remove DIVA
module 1
Anesth. agent from DIVA
module 1 may get to the sys-
tem.
Remove DIVA module 1.
Call DrägerService.
Warning 030 Remove DIVA
module 2
Anesth. agent from DIVA
module 2 may get to the sys-
tem.
Remove DIVA module 2.
Call DrägerService.
Classifi-
cation
Pri-
ority
Alarm text Cause Remedy