Instructions for use Zeus Infinity Empowered SW 2.n 433
Technical data
Radiated radio fre-
quency interference
(IEC 61000-4-3)
80 MHz to 2.5 GHz:
10 V/m
10 V/m Recommended minimum
distance from portable and
mobile radio devices with
transmitting power P
EIRP
to the medical device in-
cluding its cables:
1.84 m x PEIRP [watts]
1)
(6.04 ft x PEIRP [watts])
Conducted radio fre-
quency (IEC 61000-4-6)
150 kHz to 80 MHz:
10 V within the ISM
bands
2)
10 V Recommended minimum
distance from portable and
mobile radio devices with
transmitting power P
EIRP
to the medical device in-
cluding its cables:
(1.84 m x PEIRP [watts])
1)
(6.04 ft x PEIRP [watts])
150 kHz to 80 MHz: 3 V
outside the ISM bands
2)
3V
1) For PEIRP, insert the highest possible "equivalent isotropic radiated power" of the adjacent radio frequency transmitter.
High-frequency transmission may also occur in the vicinity of devices with the symbol . Field strengths from fixed, port-
able, or mobile radio frequency transmitters at the location of the medical device should be less than 3 V/m in the frequency
range from 150 kHz to 2.5 GHz and less than 1 V/m above 2.5 GHz.
2) ISM bands in this frequency range are: 6.765 MHz to 6.795 MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz;
40.66 MHz to 40.70 MHz.
Immunity against IEC 60601-1-2 Test
level
Compliance level of
the medical device
Electromagnetic envi-
ronment