Startup
98
Instructions for use Zeus Infinity Empowered SW 2.n
Terminating the self test:
z Touch the Cancel button (H).
If the self test is canceled during the test sequence,
the canceled message appears instead of com-
pleted.
If the self test failed, the Test details dialog window
remains open.
There is an audible tone at the end of the self test.
The system standby screen displays the following
information (I):
Exchange water traps at least every
4 weeks.
Last self test date and time
Run self test at least every 24 hours
Last test result (with colored background).
Test of the alarm system
The alarm system is automatically checked during
the self test. Additionally, the loudspeaker is auto-
matically checked when a new case is started.