12 Disturbance Analyzer Measurement
Meas Setup
30 MHZ
Sets the Frequency to 30 MHz.
Key Path
Meas Setup, Setup Table, Frequency
Example DAN:FREQ F30MHZ
Readback Text 30 MHZ
Initial S/W Revision A.14.00
Manual
Enables you to set the Frequency.
Key Path
Meas Setup, Setup Table, Frequency
Example DAN:FREQ MAN
Readback Text <freq>
Initial S/W Revision A.14.00
Key Path
Meas Setup, Setup Table, Frequency
Remote Command
[:SENSe]:DANalyzer:FREQuency:MANual <freq>
[:SENSe]:DANalyzer:FREQuency:MANual?
Example DAN:FREQ:MAN 150 KHz
DAN:FREQ:MAN?
Preset 150 KHz
State Saved Saved in instrument state
Min Depends on the instrument minimum frequency.
Max Depends on the instrument maximum frequency.
Default Unit Hz
Initial S/W Revision A.14.00
Terminal
Selects the terminal type of the CISPR limit.
Key Path
Meas Setup, Setup Table
Remote Command
CALCulate:DANalyzer:LIMit:CISPr:TERMinal MAINS | LOAD | MP1 | MP2 | MP3
CALCulate:DANalyzer:LIMit:CISPr:TERMinal?
Example CALC:DAN:LIM:CISP:TERM MAINS
CALC:DAN:LIM:CISP:TERM?
1066 EMI Receiver Mode Reference