8 Frequency Scan Measurement
Save
Dwell Time,0.00410213333333333,0.000108066666666717,6.73333333333333E–06,6.73333333336191E–
06,6.73333333335395E–06,1.26666666667647E–06,0.001,0.001,0.001,0.001
Dwell Time Mode,AUTO,AUTO,AUTO,AUTO,AUTO,AUTO,AUTO,AUTO,AUTO,AUTO
Step
Size,100,4499.5477841423,60000,60003.4287673581,60002.4743288383,500000,100000,100000,100000,100000
Step Size Mode,AUTO,AUTO,AUTO,AUTO,AUTO,AUTO,AUTO,AUTO,AUTO,AUTO
Scan Points,1411,6635,4501,11667,16167,34001,2,2,2,2
Scan Points Mode,AUTO,AUTO,AUTO,AUTO,AUTO,AUTO,AUTO,AUTO,AUTO,AUTO
Scan
Time,5.784008,0.716914266667,0.0303,0.078551066667,0.108851066667,0.043066666667,0.001,0.001,0.001,0.001
Scan Time Mode,AUTO,AUTO,AUTO,AUTO,AUTO,AUTO,AUTO,AUTO,AUTO,AUTO
Auto Step Size Rules,Points per RBW,Points per RBW,Points per RBW,Points per RBW,Points per RBW,Points per RBW,Points per
RBW,Points per RBW,Points per RBW,Points per RBW
Pts/RBW,2,2,2,2,2,2,2,2,2,2
Log %,10,10,10,10,10,10,10,10,10,10
Atten,10,10,10,10,10,10,10,10,10,10
Int Preamp,OFF,OFF,OFF,OFF,OFF,OFF,OFF,OFF,OFF,OFF
RF Input,Input1,Input1,Input1,Input1,Input1,Input1,Input1,Input1,Input1,Input1
Meas Results
Pressing this key selects Meas Result as the data type to be exported with a save request. The next step is
to select the Save As key in the Save Data menu. You are given the flexibility to choose what to display in
the report.
Key Path
Save, Data
Remote Command
:MMEMory:STORe:RESults <filename>
Example :MMEM:STOR:RES “C:\My Documents\EMI\data\report1.html”
exports the measurement result to the file report1.html
Notes The default path is at “My Documents\EMI\data\FSCAN\Results\”
If the save is initiated via SCPI, and the file already exists, the file and the directory will be
overwritten.
Using the C: drive is strongly discouraged, since it runs the risk of being overwritten during an
instrument software upgrade.
Both single and double quotes are supported for any filename parameter over SCPI.
Initial S/W Revision A.10.00
Trace
Enables you to customize the trace related information to be added into the report.
Key Path
Save, Data, Meas Result
Initial S/W Revision A.10.00
EMI Receiver Mode Reference 557

Table of Contents

Other manuals for Keysight PXA X Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight PXA X Series and is the answer not in the manual?

Keysight PXA X Series Specifications

General IconGeneral
BrandKeysight
ModelPXA X Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals