12 Disturbance Analyzer Measurement
SPAN X Scale
SPAN X Scale
Displays the menu keys that enable you to control the span x scale parameters.
Key Path
Front-panel key
Notes Front panel only.
Initial S/W Revision A.14.00
Ref Value
Specifies the Ref Value of the graph used to configure the X-axis display range.
Key Path
SPAN X Scale
Remote Command
:DISPlay:DANalyzer:VIEW[1]:WINDow[1]:TRACe:X[:SCALe]:RLEVel <time>
:DISPlay:DANalyzer:VIEW[1]:WINDow[1]:TRACe:X[:SCALe]:RLEVel?
Example DISP:DAN:VIEW:WIND:TRAC:X:RLEV 100 s
DISP:DAN:VIEW:WIND:TRAC:X:RLEV?
Couplings No
Preset 0 s
State Saved Saved in instrument state
Min 0 s
Max 7199.8 s
Initial S/W Revision A.14.00
Scale/Div
Specifies the Scale/Div value of the graph used to configure the X-axis display range.
Key Path
SPAN X Scale
Remote Command
:DISPlay:DANalyzer:VIEW[1]:WINDow[1]:TRACe:X[:SCALe]:PDIVision <time>
:DISPlay:DANalyzer:VIEW[1]:WINDow[1]:TRACe:X[:SCALe]:PDIVision?
Example DISP:DAN:VIEW:WIND:TRAC:X:PDIV 1s
DISP:DAN:VIEW:WIND:TRAC:X:PDIV?
Preset 1 s
State Saved Saved in instrument state
Min 20 ms
Max 720 s
Initial S/W Revision A.14.00