10 Monitor Spectrum Measurement
Meas Setup
Delete Unmarked
Deletes all unmarked signals in the signal list. If you accessedthis function from the front panel, the
following message will prompt you for confirmation to delete the selected signals from the signal list.
This will delete the unmarked signals. Are you sure you want to do this?
Please press Enter or OK to proceed. Press ESC or Cancel to close this dialog.
Key Path
Meas Setup, Signal List, Delete Signals
Remote Command
:CALCulate:SLISt:DELete:UNMarked
Example CALC:SLIS:DEL:UNM
Notes If the signal list contains no data, an error is generated and this function cannot be performed.
When signals in the signal list were successfully deleted, an advisory is generated.
Initial S/W Revision A.13.00
Sort Signals
Specifies how the elements in the signal list will be sorted either based on the frequency, detector
amplitude, detector vs. limit delta or time in ascending or descending order. During sorting, an advisory
message appears at the message bar, “Sorting signal list by <sorting type selected>…”, for example:
Sorting signal list by Frequency…
Sorting signal list by Det 1 Amplitude…
Sorting signal list by Det 3 vs Limit Delta…
Sorting signal list by Det 3 vs TimeStamp…
Key Path
Meas Setup, Signal List, Sort Signals
Remote Command
:CALCulate:SLISt:SORT:TYPE FREQuency|DAMPlitude|DLDelta|TIME
:CALCulate:SLISt:SORT:TYPE?
Example CALC:SLIS:SORT:TYPE FREQ
Dependencies Selecting a sorting mode (pressing the key or sending the equivalent SCPI command) will turn on the
Auto Sort state, even if that sorting mode was already selected.
Preset FREQuency
State Saved Saved in instrument state
Range By Freq|By Detector Amptd|By ( Detector|By Time
Initial S/W Revision A.13.00
By Freq
Specifies that the elements in the signal list will be sorted based on the frequency.
Key Path
Meas Setup, Signal List, Signal List, Sort Signals
786 EMI Receiver Mode Reference