3 Programming the Analyzer
STATus Subsystem
Mode All
Remote Command
:STATus:QUEStionable:INTegrity:SIGNal:PTRansition <integer>
:STATus:QUEStionable:INTegrity:SIGNal:PTRansition?
Example STAT:QUES:INT:SIGN:PTR 4 Burst not found will be reported to the Integrity Summary of the Status
Questionable register.
Preset 32767
Min 0
Max 32767
Status Bits/OPC
dependencies
Sequential command
Initial S/W Revision Prior to A.02.00
Questionable Integrity Uncalibrated Register
"Questionable Integrity Uncalibrated Condition " on page 132
"Questionable Integrity Uncalibrated Enable " on page 132
"Questionable Integrity Uncalibrated Event Query " on page 133
"Questionable Integrity Uncalibrated Negative Transition " on page 133
"Questionable Integrity Uncalibrated Positive Transition " on page 134
Questionable Integrity Uncalibrated Condition
This query returns the decimal value of the sum of the bits in the Questionable Integrity Uncalibrated
Condition register.
The data in this register is continuously updated and reflects the current conditions.
Mode All
Remote Command
:STATus:QUEStionable:INTegrity:UNCalibrated:CONDition?
Example STAT:QUES:INT:UNC:COND?
Preset 0
Status Bits/OPC
dependencies
Sequential command
Initial S/W Revision Prior to A.02.00
Questionable Integrity Uncalibrated Enable
This command determines which bits in the Questionable Integrity Uncalibrated Condition Register will set
bits in the Questionable Integrity Uncalibrated Event register, which also sets the Data Uncalibrated
Summary bit (bit 3) in the Questionable Integrity Register. The variable <integer> is the sum of the decimal
values of the bits you want to enable.
132 EMI Receiver Mode Reference