6 System Functions
System
Configuration Application Memory (Remote Command Only)
This remote command is used to query the amount of Virtual Memory a particular application consumes.
Remote Command
:SYSTem:PON:APPLication:VMEMory:USED:NAME? <INSTrument:SELect name>
Example :SYST:PON:APPL:VMEM:USED:NAME? CDMA2K
Notes <INSTrument:SELect name> is from the enums of the :INSTrument:SELect command
Value returned will be 0 (zero) if the name provided is invalid.
Preset Not affected by Preset
Initial S/W Revision Prior to A.02.00
Alignments
The Alignments Menu controls and displays the automatic alignment of the instrument, and provides the
ability to restore the default alignment values.
The current setting of the alignment system is displayed in the system Settings Panel along the top of the
display, including a warning icon for conditions that may cause specifications to be impacted.
Key Path
System
Initial S/W Revision Prior to A.02.00
Auto Align
Configures the method for which the automatic background alignment is run.
Automatic background alignments are run periodically between measurement acquisitions. The
instrument’s software determines when alignments are to be performed to maintain warranted operation.
The recommended setting for Auto Align is Normal.
An Auto Align execution cannot be aborted with the Cancel (ESC) key. To interrupt an Auto Align execution,
select Auto Align Off.
Key Path
System, Alignments
Mode All
Remote Command
:CALibration:AUTO ON|PARTial|OFF
:CALibration:AUTO?
Example :CAL:AUTO ON
Notes While Auto Align is executing, bit 0 of Status Operation register is set.
Couplings Auto Align is set to Off if Restore Align Data is invoked.
290 EMI Receiver Mode Reference