EasyManuals Logo

Keysight PXA X Series User Manual

Keysight PXA X Series
1138 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #275 background imageLoading...
Page #275 background image
6 System Functions
System
Key Path
System, Show, Errors
Mode All
Remote Command
:SYSTem:ERRor:VERBose OFF|ON|0|1
:SYSTem:ERRor:VERBose?
Example :SYST:ERR:VERB ON
Preset This is unaffected by Preset but is set to OFF on a Restore System Defaults->Misc”
State Saved No
Range On | Off
Initial S/W Revision Prior to A.02.00
Refresh
When pressed, refreshes the Show Errors display.
Key Path
System, Show, Errors
Initial S/W Revision Prior to A.02.00
Clear Error Queue
This clears all errors in all error queues.
Note the following:
Clear Error Queue does not affect the current status conditions.
Mode Preset does not clear the error queue.
Restore System Defaults will clear all error queues.
*CLS only clears the queue if it is sent remotely and *RST does not affect any error queue.
Switching modes does not affect any error queues.
Key Path
System, Show, Errors
Initial S/W Revision Prior to A.02.00
Status
See "History" on page 274.
Input Overload Enable (Remote Command Only)
Input Overload errors are reported using the Input Overload status bit (bit 12 in the Measurement Integrity
status register). Input Overloads (for example, ADC Overload errors) can come and go with great
frequency, generating many error events (for example, for signals just on the verge of overload), and so are
EMI Receiver Mode Reference 275

Table of Contents

Other manuals for Keysight PXA X Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight PXA X Series and is the answer not in the manual?

Keysight PXA X Series Specifications

General IconGeneral
BrandKeysight
ModelPXA X Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals