12 Disturbance Analyzer Measurement
Meas Setup
Preset MAINS
State Saved Saved in instrument state
Range Mains | Load | Motor Power 1 <= 700 W | Motor Power 2 700 W < MP <= 1000 W | Motor Power 3
> 1000 W
Initial S/W Revision A.14.00
Limit
Enables you to set the desired limit.
Key Path
Meas Setup, Setup Table
Remote Command
CALCulate:DANalyzer:LIMit <amptd>
CALCulate:DANalyzer:LIMit?
CALCulate:DANalyzer:LIMit:AUTO ON|OFF|1|0
CALCulate:DANalyzer:LIMit:AUTO?
Example CALC:DAN:LIM 50 dBuV
CALC:DAN:LIM?
CALC:DAN:LIM:AUTO OFF
Preset 66
ON
State Saved Saved in instrument state
Min 6.99 dBuV
Max 156.99 dBuV
Initial S/W Revision A.14.00
Attenuation
Enables you to set the value of the Attenuation parameter. This key only affects the Mechanical
Attenuator.
The following amplitude parameters are not settableby the user and therefore do not appear in any
menus:
Elec Attenuator Disabled
Elec Attenuation 0dB
Meas Atten Step 2dB
Max Mixer Level –10dBm
EMI Receiver Mode Reference 1067

Table of Contents

Other manuals for Keysight PXA X Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight PXA X Series and is the answer not in the manual?

Keysight PXA X Series Specifications

General IconGeneral
BrandKeysight
ModelPXA X Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals