Page #857 background image
10 Monitor Spectrum Measurement
Save
indicates the total amount of storage available, also in bytes. The <file_entry> is a string. Each <file_
entry> indicates the name, type, and size of one file in the directory list:
<file_name>,<file_type>,<file_size>
As the windows file system has an extension that indicates file type, <file_type> is always empty.
<file_size> provides the size of the file in bytes. For directories, <file_entry> is surrounded by square
brackets and both <file_type> and <file_size> are empty
Initial S/W Revision Prior to A.02.00
Mass Storage Change Directory (Remote Command Only)
Key path
SCPI Only
Remote Command
:MMEMory:CDIRectory [<directory_name>]
:MMEMory:CDIRectory?
Notes The string must be a valid logical path.
Changes the default directory for a mass memory file system. The <directory_name> parameter is a
string. If no parameter is specified, the directory is set to the *RST value.
At *RST, this value is set to the default user data storage area, that is defined as
System.Environment.SpecialFolder.Personal.
Query returns full path of the default directory.
Initial S/W Revision Prior to A.02.00
Mass Storage Copy (Remote Command Only)
Key path
SCPI Only
Remote Command
:MMEMory:COPY <string>,<string>[,<string>,<string>]
Notes The string must be a valid logical path.
Copies an existing file to a new file or an existing directory to a new directory.
Two forms of parameters are allowed. The first form has two parameters. In this form, the first
parameter specifies the source, and the second parameter specifies the destination.
The second form has four parameters. In this form, the first and third parameters specify the source.
The second and fourth parameters specify the directories. The first pair of parameters specifies the
source. The second pair specifies the destination. An error is generated if the source doesn’t exist or
the destination file already exists.
This command will generate an access denied” error if the destination is a restricted folder (e.g.,
C:\Windows) and the current user does not have Power User or Administrator privileges.
Mass Storage Device Copy (Remote Command Only)
This command transfers data to/from a file and a peripheral device.
EMI Receiver Mode Reference 857

Table of Contents

Other manuals for Keysight PXA X Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight PXA X Series and is the answer not in the manual?

Keysight PXA X Series Specifications

General IconGeneral
BrandKeysight
ModelPXA X Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals